Main technical parameters of 95 bit automatic purging and trapping device

source:中儀宇盛   date:2022-04-24 11:25:30   click:
  


The 95 position full-automatic purging and trapping device is a new type of full-automatic purging and trapping device of Beijing Zhongyi Yusheng Technology Co., Ltd. It can be loaded into a 95 position sample bottle once fully loaded, and can be used with all domestic and imported GC and GCMS.

It is suitable for the analysis of volatile organic compounds in liquids and solids, and can also be used for derivative reaction treatment of samples. Liquid sample, sampling to U-tube for purging; For low content solid samples, use 40ml sample bottle to purge directly; For high content solid samples, purge after automatic methanol extraction and dilution.

Types of samples that can be tested: volatile organic compounds in water, wastewater, drinking water, seawater, blood, sediment, slime, soil, cosmetics, textiles, packaging, food, spices, etc.

Main technical parameters

1. Temperature control range of sample tube:

5 ℃ to 80 ℃ above room temperature, set in increments of 1 ℃;

2. Temperature control range of sample cup:

5 ℃ to 80 ℃ above room temperature, set in increments of 1 ℃;

3. Injection needle temperature control range:

5 ℃ to 80 ℃ above room temperature, set in increments of 1 ℃;

4. Temperature control range of GC transmission pipeline:

5 ℃ to 250 ℃ above room temperature, set in increments of 1 ℃;

5. Temperature control range of valve box:

5 ℃ to 250 ℃ above room temperature, set in increments of 1 ℃;

6. Temperature control range of U-shaped purging pipe:

5 ℃ to 80 ℃ above room temperature, set in increments of 1 ℃;

7. Temperature control range of water separator:

5 ℃ to 200 ℃ above room temperature, set in increments of 1 ℃;

8. Temperature control range of pure cleaning water:

5 ℃ to 90 ℃ above room temperature, set in increments of 1 ℃;

9. Trap temperature control range:

150 ℃ to 80 ℃, set in increments of 1 ℃;

10. Trap heating rate:

>2000 ℃ / min, no more than 5% internal overshoot (at 250 ℃);

11. High temperature cooling rate of trap:

Cooling from 250 ℃ to 40 ℃ within 70 seconds;

12. Sample injection accuracy:

RSD < 1% (n = 7), determined by gravimetric method;

13. Selection of sample dilution ratio:

Programmable dilution: 1:100, 1:50, 1:25, 1:10, 1:5, 1:2;

14. Gas flow control range:

5ml / min to 500ml / min;

15. Standard 95 position sample tray (40ml sample bottle) is equipped with sample tray cooling circulating water system to realize low-temperature cooling of sample bottle;


About Us
Company Profile
Appointment interview
corporate culture
Video Center
News
Company dynamics
Common problem
Product
PURGE AND TRAP SERIES
Headspace sampler
Thermal desorption
Gas generator
Consumables and accessories
Contact us
Add:Building 13,Zhongnan High-tech Electronic InformationIndustrial Park,Yanjiao Sanhe Langfang,Hebei,China.
EMAIL:info@bjzyyskj.com.cn
Tel:+86-010-61598028 18857165617 Ms. Yang
Fax:010-61598028-108
WeChat

Scan
Copyright ? 2018 Beijing Zhongyi Yusheng Technology Co., Ltd 京ICP備10046828號-2
主站蜘蛛池模板: 一本久久道| 性感美女视频免费网站午夜 | 欧美日韩一区二区三区在线 | 日日爽爽 | 亚洲干综合 | 99国产精品久久久久久久成人热 | 五月天激情婷婷大综合 | 亚洲精品综合一区二区三区 | 日本一级毛片免费播放 | 亚洲精品一区二区三区婷婷月 | 精品国产999 | 日本天天操| 91在线视频观看 | 亚洲成人福利在线观看 | 高清人人天天夜夜曰狠狠狠狠 | 一夫三女乱爱在线播放 | 在线看片亚洲 | 欧美成人v视频免费看 | 久久成人午夜 | 激情在线日韩视频免费 | 曰本毛片 | aaa免费视频| 男人天堂网www | 日日夜夜天天操 | 日韩 亚洲 中文 图片 小说 | 欧美12一13高清视频 | 日本小视频在线 | 九色亚洲 | 一本一道波多野结衣一区二区 | 国产精品久久久久久久久久妇女 | 乱码亚洲一区二区三区 | 国产精品自在线天天看片 | 91在线一区二区三区 | 日韩在线播放一区 | 亚洲一区二区三区精品影院 | 在线视频网址免费播放 | 欧美精品午夜久久久伊人 | 色婷婷综合在线视频最新 | 乱人xxx国语对白91 | 国产爱v| 国产成人亚洲精品 |